Email updates

Keep up to date with the latest news and content from Trials and BioMed Central.

Open Access Highly Accessed Research

Compliance of a cobalt chromium coronary stent alloy – the COVIS trial

Jens Hagemeister*, Frank M Baer, Robert HG Schwinger and Hans W Höpp

Author Affiliations

Department of Medicine III, University of Cologne, Kerpener Str. 62, 50924 Cologne, Germany

For all author emails, please log on.

Current Controlled Trials in Cardiovascular Medicine 2005, 6:17  doi:10.1186/1468-6708-6-17

Published: 28 October 2005

Abstract

Background

Cobalt chromium coronary stents are increasingly being used in percutaneous coronary interventions. There are, however, no reliable data about the characteristics of unfolding and visibility of this stent alloy in vivo. The aim of this study is to compare cobalt chromium coronary stents with conventional stainless steel stents using intracoronary ultrasound.

Methods

Twenty de novo native coronary stenoses ≤ 20 mm in length (target vessel reference diameter ≥ 2.5 and ≤ 4.0 mm) received under sequential intracoronary ultrasound either a cobalt chromium stent (Multi-Link Vision®; n = 10) or a stainless steel stent (Multi-Link Zeta®; n = 10).

Results

For optimal unfolding, the cobalt chromium stent requires a higher balloon deployment pressure (13.90 ± 2.03 atm) than the stainless steel stent (11.50 ± 2.12 atm). Furthermore, the achieved target vessel diameter of the cobalt chromium stent (Visibility-Index QCA/IVUS Multi-Link Vision®1.13 / Multi-Link Zeta® 1.04) is more easily overrated by Quantitative Coronary Analysis.

Conclusion

These data indicate that stent material-specific recommendations for optimal implantation pressure and different stent material with an equal design should both be considered in interpreting QCA-analysis.